| FMPC Specifications |
| 2 Size
Channels |
0.3µm;
0.5µm |
| Flow Rate |
0.1 CFM
(2.83 L/min) |
| Light
Source |
Laser
Diode; Index Guided |
| Vacuum
Source |
Critical
Orfice requires > 18 in. Hg
vacuum |
| Inlet
Pressure |
Ambient
to 0.1 Hg vacuum |
| MBTF at
25°C Ambient |
>36,000
hours |
| Calibration |
PSL
particles in air (NIST Traceable) |
| Counting
Efficiency |
50% @
0.3µm; 100% for particles > 0.45 µm
(per JIS B9921) |
| Zero
Count |
<
1 count/5 minute (per JIS B9921) |
| Coincidence
Loss |
5% at
2,000,000 particles per ft3 |
| Interface |
RS-485
via RJ-45, Pulse mode |
| Indicators |
Power and
Calibration LED's |
| Dimensions |
4.192
(l) x 2.663 (h) x 1.618 (w) |
| Weight |
Approx.
12oz. |
| Power Requirements |
9-28 VDC (100mA
@24V max.) |
| |
|
| Operating
Conditions |
|
 |
| Relative
Humidity |
20% to
95% relative humidity, non-condensing |
| Temperature |
10°C to
40°C (50°F to 104°F) |
| |
|
| Storage
Conditions |
|
 |
| Relative
Humidity |
Up to 98%
relative humidity, non-condensing |
| Temperature |
-10°C to
50°C (14°F to 122°F) |
| |
|
| Standards |
JIS B9921 |